Human Platelet Derived Growth Factor Receptor Beta (PDGFRb) ELISA Kit

DL-PDGFRb-Hu

Overview:
Other names:CD140b; JTK12; PDGF-R-Beta; PDGFR; PDGFR1; CD140 antigen-like family member B; Platelet-derived growth factor receptor 1; Beta-type platelet-derived growth factor

Function: Receptor that binds specifically to PDGFB and PDGFD and has a tyrosine-protein kinase activity. Phosphorylates Tyr residues at the C-terminus of PTPN11 creating a binding site for the SH2 domain of GRB2.

Sequence:
MRLPGAMPAL  ALKGELLLLS  LLLLLEPQIS  QGLVVTPPGP  ELVLNVSSTF  
VLTCSGSAPV  VWERMSQEPP  QEMAKAQDGT  FSSVLTLTNL  TGLDTGEYFC  
THNDSRGLET  DERKRLYIFV  PDPTVGFLPN  DAEELFIFLT  EITEITIPCR  
VTDPQLVVTL  HEKKGDVALP  VPYDHQRGFS  GIFEDRSYIC  KTTIGDREVD 
SDAYYVYRLQ  VSSINVSVNA  VQTVVRQGEN  ITLMCIVIGN  EVVNFEWTYP  
RKESGRLVEP  VTDFLLDMPY  HIRSILHIPS  AELEDSGTYT  CNVTESVNDH  
QDEKAINITV  VESGYVRLLG  EVGTLQFAEL  HRSRTLQVVF  EAYPPPTVLW  
FKDNRTLGDS  SAGEIALSTR  NVSETRYVSE  LTLVRVKVAE  AGHYTMRAFH  
EDAEVQLSFQ  LQINVPVRVL  ELSESHPDSG  EQTVRCRGRG  MPQPNIIWSA  
CRDLKRCPRE  LPPTLLGNSS  EEESQLETNV  TYWEEEQEFE  VVSTLRLQHV  
DRPLSVRCTL  RNAVGQDTQE  VIVVPHSLPF  KVVVISAILA  LVVLTIISLI  
ILIMLWQKKP  RYEIRWKVIE  SVSSDGHEYI  YVDPMQLPYD  STWELPRDQL  
VLGRTLGSGA  FGQVVEATAH  GLSHSQATMK  VAVKMLKSTA  RSSEKQALMS  
ELKIMSHLGP  HLNVVNLLGA  CTKGGPIYII  TEYCRYGDLV  DYLHRNKHTF  
LQHHSDKRRP  PSAELYSNAL  PVGLPLPSHV  SLTGESDGGY  MDMSKDESVD  
YVPMLDMKGD  VKYADIESSN  YMAPYDNYVP  SAPERTCRAT  LINESPVLSY  
MDLVGFSYQV  ANGMEFLASK  NCVHRDLAAR  NVLICEGKLV  KICDFGLARD  
IMRDSNYISK  GSTFLPLKWM  APESIFNSLY  TTLSDVWSFG  ILLWEIFTLG  
GTPYPELPMN  EQFYNAIKRG  YRMAQPAHAS  DEIYEIMQKC  WEEKFEIRPP  
FSQLVLLLER  LLGEGYKKKY  QQVDEEFLRS  DHPAILRSQA  RLPGFHGLRS  
PLDTSSVLYT  AVQPNEGDND  YIIPLPDPKP  EVADEGPLEG  SPSLASSTLN  
EVNTSSTISC  DSPLEPQDEP  EPEPQLELQV  EPEPELEQLP  DSGCPAPRAE  
AEDSFL
2, Features

DL-PDGFRb-Hu

Human Platelet Derived Growth Factor Receptor Beta (PDGFRb) ELISA Kit
INTENDED USE
The kit is a sandwich enzyme immunoassay for the in vitro quantitative measurement of PDGFRb in human tissue homogenates, cell lysates and other biological fluids.

DETECTION RANGE
0.156-10ng/mL. The standard curve concentrations used for the ELISA’s were 10ng/mL, 5ng/mL, 2.5ng/mL, 1.25ng/mL, 0.625ng/mL, 0.312ng/mL, 0.156ng/mL.

SENSITIVITY
The minimum detectable dose of PDGFRb is typically less than 0.063ng/mL.
The sensitivity of this assay, or Lower Limit of Detection (LLD) was defined as the lowest protein concentration that could be differentiated from zero. It was determined by adding two standard deviations to the mean optical density value of twenty zero standard replicates and calculating the corresponding concentration.

SPECIFICITY
This assay has high sensitivity and excellent specificity for detection of PDGFRb.
No significant cross-reactivity or interference between PDGFRb and analogues was observed.
Note:
Limited by current skills and knowledge, it is impossible to perform all possible cross-reactivity detection tests between PDGFRb and all analogues, therefore, cross reactivity may still exist.

PRECISION
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level PDGFRb were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level PDGFRb were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

STABILITY
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions.
Note:
To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity, and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.

ASSAY PROCEDURE SUMMARY
1.Prepare all reagents, samples and standards;
2.Add 100µL standard or sample to each well. Incubate 2 hours at 37oC;
3.Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37oC;
4.Aspirate and wash 3 times;
5.Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37oC;
6.Aspirate and wash 5 times;
7.Add 90µL Substrate Solution. Incubate 15-25 minutes at 37oC;
8.Add 50µL Stop Solution. Read at 450nm immediately.

IMPORTANT NOTES
1.Limited by the current conditions and scientific technology, it is impossible to conduct comprehensive identification and analysis tests on the raw materials provided by suppliers. As a result, it is possible there are some qualitative and/or technical risks.
2.The final experimental results will be closely related to the validity of the products, operation skills of the end users and the experimental environments. Please make sure that sufficient samples are available to obtain accurate results.
3.Kits from different batches may be a little different in detection range, sensitivity and color developing time. Please perform the experiment exactly according to the instruction manual included in your kit. Electronic ones on our website are for reference only.
4.Do not mix or substitute reagents from one kit lot to another. Use only the reagents supplied by manufacturer.
5.Protect all reagents from strong light during storage and incubation. All bottle caps of reagents should be closed tightly to prevent evaporation of liquids and contamination by microorganisms.
6.There may be a foggy substance in the wells when the plate is opened at the first time. It will not have any effect on the final assay results. Do not remove microtiter plate from the storage bag until needed.
7.Incorrect procedures during reagent preparation and loading, as well as incorrect parameter setting for the plate reader may lead to incorrect results. A microplate plate reader with a bandwidth of 10nm or less and an optical density range of 0-3 O.D. or greater at 450 ± 10nm wavelength is acceptable for use in absorbance measurement. Please read the instruction carefully and adjust the instrument prior to the experiment.
8.Even the same experimenter may get different results from two separate experiments. In order to get better reproducible results, the operation of every step in the assay should be controlled. Furthermore, a preliminary experiment before the general assay for each batch is recommended.
9.Each kit has undergone several rigorous quality control tests. However, results from end users might be inconsistent with our in-house data due to some unexpected transportation conditions or different lab equipment. Intra-assay variance among kits from different batches could arise from the above factors as well.
10.Kits from different manufacturers with the same item might produce different results, since we have not compared our products with other manufacturers.
11.The standard of the kit and immunogen used for antibody preparation are commonly recombinant proteins. Different expressed sequence, expression systems, purification methods might be used in recombinant protein preparation. Besides, there might exist differences on the screening technique of antibody and antibody pairs in our kit. Thus we can not guarantee the kit could detect recombinant protein from other companies. So, it is not recommended to use the kit for the detection of recombinant protein.
12.Validity period: 12 months.
13.The instruction manual also works with the 48T kit, but all reagents in the 48T kit are reduced by half.

You can reference link of the kit as following

http://www.dldevelop.com/Research-reagent/dl-pdgfrb-hu.html

Send your message to us:

INQUIRY NOW